2026 · Conference presentation
eXaSkin and eaSyFix at ASRT 2026
Alejandro Brussain, Anatomical Geometry. ASRT Radiation Therapy Conference, 2026.
The presentation uses clinical examples to illustrate custom eXaSkin bolus and eaSyFix positioning with Moldcare. It includes setup photographs and selected dosimetry examples.
2026 · Conference poster and published abstract
Robust dosimetry at the high-density moldable bolus edge transition despite submillimetric thickness and electronic disequilibrium
Sánchez-Esperón J, Picón-García F, Juan-Martos E, Velázquez-Miranda S. Physica Medica. 2026;149.
The authors evaluated dose at the transition of a stepped eXaSkin bolus, with thicknesses from 0.8 to 7.8 mm, using film measurements and Acuros XB. The mean gamma value was 0.13 under the study’s 3%/3 mm criterion.
2020 · EFOMP conference poster
Bolus as a shaper in the superficial lesions treatment with an electron beam
Derecho-Torres FA, López-Martín N, Velázquez-Miranda S, Luís-Simón FJ. EFOMP, 2020.
A phantom experiment examined a molded eXaSkin ring as a beam-shaping material for 6 MeV electrons. In this setup, film verification reported a 93.28% gamma pass rate at 3%/3 mm.
2019 · Published conference paper
Dose build up characteristics with eXaSkin bolus during 6 MV radiotherapy: MOSkin dosimetry results
Al-Sudani T et al. Journal of Physics: Conference Series. 2019;1154:012024.
The researchers compared eXaSkin and solid water using MOSkin and an advanced Markus chamber. The measured depth of maximum dose was about 9 mm in eXaSkin versus 15 mm in solid water in the reported 6 MV setup.
2018 · AAPM conference report
Customized bolus is more effective than standard bolus to mitigate the increase in exit dose during MRI-guided radiation therapy
David S, Lee H, Ibbott G. AAPM Annual Meeting, 2018.
The authors compared molded eXaSkin, 3D printed PLA, and Superflab for exit surface dose on three phantoms. In the MR-linac tests, the effect depended on phantom shape and whether ultrasound gel was used.
2018 · ESTRO 37 poster · EP-2174
Fabrication of three-dimensional printed customized bolus for the irregular shape of the outer ear
Baeza M et al. ESTRO 37, 2018.
A head phantom study compared a 3D printed outer-ear bolus, eXaSkin, and no bolus. The two bolus plans showed similar target coverage in the reported comparison, with some air gaps visible for both boluses.
2013 · Published conference abstract
Dosimetric characterization of eXaSkin: High density bolus for photon use
Ortiz Seidel M et al. Reports of Practical Oncology and Radiotherapy. 2013;18(Suppl 1):S355.
An early conference report on the dosimetric characterization of eXaSkin for photon radiotherapy.
2012 · ESTRO 31 poster · EP-1235
New high density bolus: A new approach to treatment of superficial tumours with photons
Ortiz M, Velázquez S. ESTRO 31, 2012.
The investigators compared measured 6 MV depth dose curves with and without a 1 cm high-density bolus and assessed agreement with treatment planning calculations.